Fluke T150 Manual

4.5 · 35 Ratings

PDF manual · 4 Pages

English

M: OA-GPT-5-NAN
page

/ 4

Loading...
44b308f67d51a3ddc44b3b03f5d30f8e626c page 4
Environmental
Pollution degree .......................2
Protection degree .....................IP54 (T90)
IP64 (T110/T130/
T150)
Operating Temperature ............-15 °C to +45 °C
Storage Temperature ................-20 °C to +60 °C
Humidity ...................................85 % RH maximum
Altitude .....................................2000 m
Vibration ...................................refer to EN61243-3
Safety EN61243-3:2010
Transporting goods...................VBG 1, § 35
Overvoltage protection .............690 V ac/dc
Measurement category
T90......................................CAT II 690 V
CAT III 600 V
T110/T130/T150 .................CAT III 690 V
CAT IV 600 V
Power supply ...............................2 x 1.5 V Micro /
LR03 / AAA
Power consumption ....................50 mA maximum /
~250 mW
Language support .......................Czech, Dutch,
English, Finnish,
French, German,
Italian, Norwegian,
Polish, Portuguese,
Romanian,
Russian, Spanish,
Swedish, Turkish
LIMITED WARRANTY & LIMITATION OF
LIABILITY
This Fluke product will be free from defects in material and
workmanship for two years from the date of purchase. This
warranty does not cover fuses, disposable batteries, or damage
from accident, neglect, misuse, alteration, contamination, or
abnormal conditions of operation or handling. Resellers are not
authorized to extend any other warranty on Fluke’s behalf. To
obtain service during the warranty period, contact your nearest
Fluke authorized service center to obtain return authorization
information, then send your defective Tester to that Service
Center with a description of the problem. Replace depleted
batteries immediately to avoid Tester damage from battery
leakage.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM
ANY CAUSE OR THEORY. Since some states or countries do
not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability
may not apply to you.
Fluke Corporation Fluke Europe B.V
P.O. Box 9090 P.O. Box 1186
Everett WA 98206‑9090 5602 B.D.
Eindhoven
The Netherlands
11/99
Specications
Model
T90 T110 T130 T150
LEDs
Voltage range 12 V to 690 V ac/dc
• • • •
Resolution
±12 V, 24 V, 50 V, 120 V,
230 V, 400 V, 690 V
• • • •
Tolerance
Complies with
EN 61243-3:2010
• • • •
Frequency range 0 / 40 Hz to 400 Hz
• • • •
Response time ≤0.5 second
• • • •
Auto power on ≥12 V ac/dc
• • • •
LCD
Voltage range 6 V to 690 V ac/dc
• •
Resolution ±1 V
• •
Tolerance ±(3 % rdg + 5 digits)
• •
Frequency range 0 / 40 Hz to 400 Hz
• •
Response time ≤1 second
• •
Auto power on ≥6 V ac/dc
• •
Voltage detection Automatic
• • • •
Polarity detection Full range
• • • •
Range detection Automatic
• • • •
Internal basic load impedance
Peak current
Maximum 3.5 mA at
690 V 200 kΩ / Is <3.5 mA
(no RCD tripping)
• • • •
Operation time
Duration Time =
30 seconds
• • • •
Recovery time
Recovery Time =
240 seconds
• • • •
Switchable Load ~7 kΩ
• • •
Peak current Is (load) = 150 mA
• • •
RCD tripping I~30 mA @ 230 V
• • •
Continuity Test 0 to 400 kΩ
• • • •
Accuracy nominal resistance +50 %
• • • •
Test current ≤5 µA
• • • •
Single-pole Phase Test 100 V ac to 690 V ac
• • • •
Frequency range
40 Hz to 60 Hz
•
50 Hz to 400 Hz
• • •
Rotary Field Indication
• • •
Voltage range (LEDs)
100 V to 690 V
(phase to earth)
• • •
Frequency range 50 Hz to 60 Hz
• • •
Resistance Measurement 0 Ω to 1999 Ω
•
Resolution 1 Ω
•
Tolerance
±(5 % rdg +10 digits) @
20 °C
•
Temperature coefcient ±5 digits / 10 K
•
Test current ≤30 µA
•
Size in mm (HxWxL) 245x64x28 255x78x35
Weight in kg (includes batteries) 0.18 0.27
How to Clean
Before you clean the Tester, remove it from all
measurement circuits.
Caution
To prevent damage, do not use abrasives or
solvents on the Tester.
Clean the case with a moist cloth and weak detergent.
After you clean the Tester, do not use it for a period
of 5 hours.
When to Calibrate
Fluke recommends a calibration interval of 1 year.
Battery Replacement
If b (Fluke T90/T110) is on or B shows
in the LCD (Fluke T130/T150) during tests or
measurements, replace the batteries.
To replace the batteries:
1. Disconnect the Tester from the measurement
circuit.
2. Open the battery cover. See Figure 8.
3. Remove the discharged batteries.
4. Replace with two new 1.5V IEC LR03 AAA
batteries.
5. Align the battery polarity as shown on the case
housing.
6. Close and attach the battery cover.
Note
Do not overtighten the screw for the battery cover.
7. Complete a self‑test.
8

Chat with your Fluke T150 manual using AI

Get instant AI answers from your Fluke T150 manual

Ask anything about the Fluke T150. Answers come from this manual, with the pages they are based on.

T150 Maintenance Schedule

Remind me when maintenance is due
Yearly
Calibrate the tester

Fluke recommends a calibration interval of 1 year.

Before and after each use
Perform the tester self-test

Touch probe tips together, verify battery and indicators, then measure a known voltage. Do not use the tester if it fails.

Before each use
Inspect the tester case

Examine the case for cracks or missing plastic. Do not use the product if it is damaged.

Free maintenance reminders

Timed to your purchase date. Always free, unsubscribe anytime.

By submitting you agree to the privacy policy. Unsubscribe with one click in any email.
M: gpt-5.6-terra

Ask a Question

0/1000

Do you recommend T150?

Yes No
0
User Pic

Laura Schmidt

10 months ago

Can I switch off the annoying beep or acoustic signal on the Fluke T150/VDE when verifying mains voltage, or is it always on?

0
User Pic

Manual.ly

10 months ago

On the Fluke T150/VDE, the acoustic signal is intended as a safety alert and, in this version, it cannot be fully disabled. Users therefore hear the tone during voltage verification and during certain test functions. To reduce distraction, rely on the visual indications on the display/LEDs and the measurement light instead of the sound. If your unit ever behaves unusually, contact support with the device details so it can be checked or replaced under warranty.

M: OA-GPT-5-NAN
M: OA-GPT-5-NAN

Video Manual

Fluke Spannungs- und Durchgangsprüfer T150/VDE Vorstellung + Funktionserläuterung

Frequently Asked Questions

A

The T150 detects voltages from low levels upward: LEDs operate from 12 V to 690 V AC/DC; the LCD displays from 6 V to 690 V with 1 V resolution on models with an LCD.

Is this answer helpful?

A

Yes. By pressing both load buttons simultaneously you activate a low-impedance mode that can trip typical 10 mA or 30 mA RCDs in a 230 V system. The load is overload-protected and will reduce current after ~20 seconds at 230 V (or ~2 seconds at 690 V).

Is this answer helpful?

A

The single-pole phase test identifies phase presence by holding the indicator probe body and touching the tip to a conductor; it signals above ~100 V AC. It’s meant for quick checks but not to validate zero voltage—use the voltage measurement mode for precise verification.

Is this answer helpful?

A

The T150 offers CAT III/CAT IV overvoltage ratings, double insulation, fixed probe cover, finger guards, an independent ELV LED (>50 V AC or >120 V DC), IP64 protection, and built-in self-tests to confirm correct operation before use.

Is this answer helpful?

A

Yes. The T150 measures resistance from 0 to 1999 Ω with 1 Ω resolution; typical tolerance is ±(5 % reading + 10 digits) at 20 °C. The resistance function must be enabled and the unit should be used only on de-energized circuits.

Is this answer helpful?

A

Remove batteries when storing long-term, replace batteries when the low-battery indicator appears, align polarity when installing, and perform a self-test after replacement. Promptly address any battery leakage and have repairs performed by approved technicians.

Is this answer helpful?

A

The beeper is switchable on models with the feature. Press and hold the beeper button for two seconds to toggle it on or off. Always test continuity with probes touched together to confirm the beeper functions before use.

Is this answer helpful?

M: OA-GPT-5-NAN

Advantages and Disadvantages

Advantages

  • High safety ratings (CAT III 690 V / CAT IV 600 V) and IP64 protection for rugged, on-site use.
  • Switchable low-impedance load allows RCD trip testing and reduces interference from stray coupling.
  • Compact slim probes, probe-tip storage cap, and accessory sheaths for tight spaces and organized transport.
  • Multi-function: voltage, continuity, diode, resistance (1–1999 Ω), single-pole phase, and rotary field indication.
  • Built-in self-test, independent ELV LED, and beeper/hold/torch features improve reliability and usability.
VS

Disadvantages

  • LCD numeric measurement is available only on advanced variants (T130/T150), limiting exact readings on base models.
  • Switchable load is time-limited by overload protection—current reduces after ~20s at 230 V and ~2s at 690 V.
  • Resistance measurement range is limited to 0–1999 Ω, unsuitable for high-resistance insulation testing.
  • Beeper and some functions may be less effective in high ambient noise; always verify audibly before tests.
  • Requires AAA batteries; battery leakage can damage the unit if left unused without removal.
M: OA-GPT-5-NAN

Product Specification

Measurement Range

6-690V

M: OA-GPT-54-NAN

Product Description

Fluke T150 VDE Voltage Tester — Reliable, Compact, and Safe

The Fluke T150 VDE Voltage Tester is a compact, rugged voltage and continuity tester designed for electricians and technicians working in industrial, commercial, and residential environments. Combining a clear LED bargraph with an LCD readout (on advanced models) and a range of safety-focused features, the T150 delivers fast visual indication of voltages from low-voltage circuits up to high mains levels. Its slim probe tips, fixed probe cover, and integrated probe-tip protector cap make it convenient for tight spaces and safe storage.

Built to meet strict safety standards, the T150 features CAT III/CAT IV overvoltage protection ratings, IP64 ingress protection, and double-insulated wiring. The unit offers automatic power-on detection above low-voltage thresholds, a visual >ELV indicator that flags voltages above 50 V AC or 120 V DC, and an internal low-impedance mode with a switchable load for RCD trip tests. These options let users reduce interference from capacitive or inductive coupling and verify RCD operation by applying realistic trip currents during measurements.

Beyond basic voltage testing, the T150 provides continuity and diode testing, a low-ohm resistance function (1–1999 Ω) with 1 Ω resolution, and a rotary field indicator for three-phase phase-sequence checks. A useful single-pole phase test and an independent ELV indicator LED ensure rapid detection of live conductors even if batteries fail. Additional conveniences include a torch/backlight for dim workspaces, a beeper (switchable) for continuity and phase indications, and a HOLD function for capturing readings on the LCD.

Ergonomically designed, the tester supports single-handed operation and can be used while wearing gloves. Built-in self-test and clear battery-low indicators help ensure dependable operation before and after use. Accessories such as probe tip sheaths, 4 mm probe extensions, and a holster/carry case improve versatility on the job.

This tester is ideal for electricians who need a portable, safety-certified instrument that handles common tasks—voltage detection, continuity checks, RCD trip testing, and phase sequencing—while providing robust protection and user-friendly features.

Notes

  • The >ELV LED warns at >50 V AC or >120 V DC and should not replace true measurement validation. Use the LCD for numeric readings when needed.
  • Switchable load (two-button press) is time-limited by overload protection; expect current reduction after ~20 seconds at 230 V.
  • Always perform the self-test and verify the beeper/continuity before starting live work.