TRITEST
®
pro
3
Operating Manual
TRITEST
®
pro
Before using the TRITEST
®
pro phase sequence
indicator:
Pleasereadtheoperatingmanualandabsolutely
observethesafetyinstructions!
Table of contents
1. Safetyinstructions
2. Device description
3. Functional test
4. Phase sequence test
5. LED torch
6. Batteryreplacement
7. Technical data
8. General maintenance
9. Environmental protection
1. Safetyinstructions
Attention! Magnets might affect the cor-
rect functioning of cardiac pacemakers
and implanted defibrillators. As a user of
such medical devices, keep a sufficient
distance to the magnet.
- During the tests, touch the device at the insulated
test probes
9
, and only and do not touch
the bare testing electrodes
8
!
- Check the device for correct functioning immedi-
ately before and after using it (see section 3)! Do
not use the device, if one or more indications are
not working or if it does not seem to be ready for
operation!
- If it is assumed that safe operation is no longer
possible, the device must be switched off immedi-
ately.
- Absolutely prevent the device from getting wet
and avoid any water condensation at the device.
Moreover, the device must be protected against
contamination and damaging!
- The device must be used within the stated nomi-
nal voltage range and in electrical installations of
up to 500 V AC only!
- The device must be used only in electric circuits
of overvoltage category CAT III with max. 300 V
for phase-to-earth measurements. For measure-
ments within measuring category III, the protrud-
ing conductive part of a testing electrode
8
of the
test probe must not be longer than 4 mm. Before
carrying out measurements within measuring cat-
egory III, the enclosed attachable protective caps
marked with CAT III must be pushed onto the test-
ing electrodes
8
for user protection purposes.
- Please observe that work on live parts and elec-
trical components of all kinds is dangerous! Even
low voltages of 30 V AC and 60 V DC may be
dangerous to human life!
- Do not operate the device with the battery com-
partment being open.
- The device is designed for being used by qualified
electricians and under safe working conditions.
- Do not dismantle the device!
Attention!
The device is provided with a high-power LED torch.
Do not look directly or indirectly via reflecting surfaces
into the LED beam. An LED beam might cause irrepa-
rable damage to your eyes.
Symbols on the device:
Symbol Meaning
Attention! Please observe documenta-
tion!
This symbol indicates that the informa-
tion provided in the operating manual
must be complied with in order to avoid
risks.
Alternating voltage (AC)
Earth (voltage to earth)
This symbol shows the orientation of the
batteries for inserting them with correct
polarity.
Warning! Potentially dangerous optical
radiation!
Do not look directly into the light beam!
Danger to your retina!
Attention! Magnets might affect the cor-
rect functioning of cardiac pacemakers
and implanted defibrillators. As a user of
such medical devices, keep a sufficient
distance to the magnet.
2. Device description
1
Red LED for external conductor (phase) L1
2
Red LED for external conductor (phase) L2
3
Red LED for external conductor (phase) L3
4
Red LED for counter-clockwise phase sequence
5
Green LED for clockwise phase sequence
6
High-power LED torch
7
Push-button
8
Testing electrodes with attachable protective caps
9
Test probe L1 (brown), part no.: 10112582
Test probe L2 (black), part no.: 10112581
Test probe L3 (grey), part no.: 10112584
Alligator clip, part no.: 709269
Catch of the battery compartment cover
Battery compartment cover with magnet, belt clip
and technical data
3. Functional test
- Check the device for correct functioning immedi-
ately before and after using it!
- Test the correct functioning of the LEDs
1
,
2
and
3
for indicating the external conductors L1,
L2 and L3 as well as the LEDs
4
and
5
of the
phase sequence indication with a familiar voltage
source, e.g. a 400 V CEE socket.
- Do not use the device, if not all functions are
working properly!
- Please replace the batteries as soon as the lumi-
nosity of the LED torch
6
decreases.
4. Phase sequence test
- Connect the connecting cables marked with L1,
L2 and L3 to the mains to be tested. For this pur-
pose, use the test probes
9
, and or the alli-
gator clip .
- The external conductor voltages are indicated by
means of the red LEDs
1
,
2
and
3
.
- Depending on the connection, the red LED (4) for
"counter-clockwise phase sequence" or the green
LED
5
for "clockwise phase sequence" will be
activated.
- Interchanging the test probes on two external
conductors must result in a change of the phase
sequence.
- For each test, make sure that all three external
conductor voltages are present. Only then, the
indicated results can be analyzed!

Chat with your Benning TRITEST pro manual using AI

Get instant AI answers from your Benning TRITEST pro manual

Ask anything about the Benning TRITEST pro. Answers come from this manual, with the pages they are based on.

TRITEST pro Maintenance Schedule

Before and after each use
Check device functionality

Test the LEDs for phase indication and rotation direction on a known voltage source immediately before and after every use.

As needed
Replace batteries

Replace the three AA (LR06) batteries when the brightness of the LED torch decreases.

Regularly / As needed
Clean device exterior

Clean the outside of the housing with a clean dry cloth. Also clean the battery compartment if contaminated.

Before long-term storage
Remove batteries

If the device is stored for a longer period of time, remove the batteries from the device to prevent damage.

M: qwen/qwen3.8-flash

Ask a Question

0/1000

Do you recommend TRITEST pro?

Yes No
0
User Pic

Wellington Larson

6 months ago

Does the Benning TRITEST pro include a flashlight and magnetic mount for easier handling?

0
User Pic

Manual.ly

6 months ago

Yes, the device features an integrated LED flashlight activated by a side button, a magnet on the back, and a belt clip to assist with positioning during use.

M: QW-35-FL
M: QW-35-FL
0
User Pic

Duane Kuvalis

5 months ago

Are the test leads color-coded and labeled for users with color blindness?

M: QW-35-FL

Frequently Asked Questions

A

Yes, the TRITEST pro phase sequence test works fully even when the batteries are removed or completely discharged. However, the LED flashlight will not function without power.

Is this answer helpful?

A

No, the device contains magnets that can affect cardiac pacemakers and implanted defibrillators. Users with these devices must maintain a safe distance from the instrument.

Is this answer helpful?

A

The device is rated for a nominal voltage range of 400 V to 500 V AC. It should only be used in circuits belonging to overvoltage category CAT III with a maximum of 300 V to earth.

Is this answer helpful?

A

Open the battery compartment on the back of the device by pressing the catch and sliding the cover down. Replace the three old AA batteries with new ones, ensuring correct polarity, then slide the cover back until it clicks.

Is this answer helpful?

A

The green LED illuminates to indicate a clockwise (right-hand) rotating field sequence. If the red LED for left-hand rotation lights up instead, the sequence is reversed.

Is this answer helpful?

A

No, the device has an IP 40 rating which protects against solid objects but offers no water protection. You must keep the device dry and free from condensation at all times.

Is this answer helpful?

M: QW-35-FL

Advantages and Disadvantages

Advantages

  • Clear visual indication of phase sequence with distinct red and green LEDs
  • Operates effectively even with removed or dead batteries
  • Includes a powerful integrated LED flashlight for low-light conditions
  • Robust build quality with IP 40 protection against dust and solids
  • Convenient belt clip and magnet for easy attachment to tools or belts
VS

Disadvantages

  • Not waterproof and offers no protection against water ingress
  • Magnets inside the case can interfere with pacemakers and defibrillators
  • Requires careful handling to avoid direct exposure to the bright LED beam
  • Limited to specific voltage ranges and measurement categories
  • Must be used by qualified personnel due to high-voltage risks
M: QW-35-FL

Product Specification

Measurement range

400-500V

M: QW-35-FL

Product Description

Benning Tritest pro Rotating Field Direction Indicator Phase Tester

The Benning Tritest pro is a specialized instrument designed for electricians to verify the phase sequence and direction of rotating fields in three-phase electrical systems. This robust device operates within a nominal voltage range of 400 V to 500 V AC, making it ideal for industrial applications where precise phase rotation checks are critical for motor safety and operation.

Equipped with intuitive LED indicators, the tester provides immediate visual feedback. Three red LEDs confirm the presence of voltage on phases L1, L2, and L3. A fourth red LED signals a counter-clockwise (left-hand) phase sequence, while a green LED indicates a clockwise (right-hand) sequence. This clear distinction helps professionals quickly identify wiring errors before connecting motors or other sensitive equipment.

Beyond its primary function, the unit features a high-power LED flashlight for use in dimly lit environments like switchgear rooms. The design prioritizes user safety with insulated test probes and specific warnings regarding magnetic interference with medical devices such as pacemakers. The device is built to withstand harsh conditions with an IP 40 rating, protecting against solid objects larger than 1 mm, though it is not waterproof.

Maintenance is straightforward, requiring only periodic battery replacement using three standard AA batteries. Interestingly, the phase sequence test remains functional even if the batteries are removed or depleted, ensuring reliability in critical situations. The unit includes a belt clip and magnet on the battery cover for convenient hands-free carrying during field work.

Key Features

  • Direct reading of phase sequence without complex calculations.
  • Integrated high-intensity LED torch for visibility.
  • Safe operation up to CAT III 300 V limits.
  • Durable construction suitable for professional trade use.

Notes

Always ensure all three phases are present before analyzing the sequence indicator. Avoid looking directly into the LED beam to prevent eye damage.