BENNING CM 1-3
9
Operating Manual
BENNING CM 1-3
Digital current clamp multimeter for
- AC current measurements
- AC voltage measurements
- DC voltage measurements
- resistance measurements
- diode tests
- continuity tests
Table of contents
1. User instructions
2. Safety instructions
3. Scope of delivery
4. Device description
5. General information
6. Ambient conditions
7. Electrical specifications
8. Measuring with the BENNING CM 1-3
9. Maintenance
10. Technical data of measuring accessories
11. Environmental note
1. User instructions
This operating manual is intended for
- skilled electricians and
- electrotechnically trained personnel.
The BENNING CM 1-3 is intended for measurements under dry ambient condi-
tions. It must not be used in electrical circuits with a nominal voltage higher
then 1000 V DC and 750 V AC (see section 6 „Ambient conditions“ for details).
The following symbols are used in this operating manual and on the
BENNING CM 1-3:
Application around and removal from HAZARDOUS LIVE conduc-
tors is permitted.
Warning of electrical danger!
Indicates instructions which must be followed to avoid danger to
persons.
Attention! Must comply with documentation!
This symbol indicates that the information provided in the operating
manual must be complied with in order to avoid risks.
This symbol on the BENNING CM 1-3 indicates that the
BENNING CM 1-3 is equipped with protective insulation (protec-
tion class II).
This symbol appears on the display to indicate a discharged
battery.
This symbol designates the „diode test“ field.
This symbol designates the „continuity test“ field.
The buzzer is intended for acoustic result output.
(DC) Direct voltage
(AC) Alternating voltage or current
Ground (voltage against ground)

Chat with your Benning CM 1-3 manual using AI

Get instant AI answers from your Benning CM 1-3 manual

Ask anything about the Benning CM 1-3. Answers come from this manual, with the pages they are based on.

CM 1-3 Maintenance Schedule

Remind me when maintenance is due
Yearly
Calibrate the device at factory service

Send the device to Benning Service Center for recalibration once a year to maintain specified measurement accuracy.

Show in the manual (page 13)
As needed
Replace batteries when low battery symbol appears

Replace the two 1.5V micro batteries immediately if the battery symbol appears on the display. Ensure device is voltage-free before opening.

Show in the manual (page 13)
Regularly
Clean housing and battery compartment

Wipe exterior with a clean dry cloth. Check battery contacts for electrolyte leakage or white deposits and clean them with a dry cloth if found.

Show in the manual (page 13)

Free maintenance reminders

Timed to your purchase date. Always free, unsubscribe anytime.

By submitting you agree to the privacy policy. Unsubscribe with one click in any email.
M: qwen/qwen3.8-flash

Ask a Question

0/1000

Do you recommend CM 1-3?

Yes No

Frequently Asked Questions

A

The device can measure up to 1000 V DC and 750 V AC. It is rated for overvoltage category III at 1000 V and category IV at 600 V relative to earth.

Is this answer helpful?

A

Select the continuity function on the rotary switch and press the HOLD key for continuity mode. Connect the leads to the points being tested. If resistance is below 50 ohms, the internal buzzer will sound.

Is this answer helpful?

A

No, the device is designed strictly for measurements in dry ambient conditions. It has an IP 30 rating which offers no protection against water.

Is this answer helpful?

A

Using alkaline batteries, the device offers an approximate battery life of 250 hours. The meter will display a battery symbol when the voltage drops below the operating threshold.

Is this answer helpful?

A

The VoltSensor button activates a non-contact voltage indicator. It detects AC fields and signals the presence of phase voltage via a red LED and an audible beep without needing test leads.

Is this answer helpful?

A

Yes, the Benning CM 1-3 automatically switches off after approximately 10 minutes of inactivity to preserve battery power.

Is this answer helpful?

M: QW-35-FL

Advantages and Disadvantages

Advantages

  • Supports multiple measurement types including AC/DC voltage, current, resistance, and diode tests
  • Includes a non-contact VoltSensor for safe phase detection and cable break tracing
  • Built-in buzzer provides immediate acoustic feedback during continuity testing
  • Complies with high safety standards (CAT III 1000V / CAT IV 600V) for professional use
  • Auto-power off feature conserves battery life after ten minutes of inactivity
VS

Disadvantages

  • Intended only for dry environments and cannot withstand water exposure
  • Requires annual calibration by service center to maintain specified accuracy
  • Battery replacement requires removing the cover and unscrewing a fastener
  • Maximum resolution limited to 1999 counts on the digital display
  • Measuring leads must be inspected regularly for insulation damage to ensure safety
M: QW-35-FL

Product Specification

Measurement range

max. 200A, max. 100V, max. 20MΩ

M: QW-35-FL

Product Description

Benning CM 1-3 Clamp Meter-Multimeter: A Comprehensive Review

The Benning CM 1-3 is a robust digital clamp meter designed for skilled electricians and electrotechnically trained personnel who require precision in dry environments. This versatile tool combines the functionality of a standard multimeter with a fixed jaw current sensor, allowing for efficient AC current measurements without breaking the circuit. The device features a clear 3½-digit liquid crystal display that presents readings up to 1999, along with automatic polarity indication and a decimal point for easy interpretation.

Equipped with a rotary switch, users can seamlessly toggle between various measurement modes including AC voltage, DC voltage, resistance, diode testing, continuity checks with an audible buzzer, and AC current. The unit operates safely within overvoltage categories III and IV, supporting up to 1000 V DC and 750 V AC. Safety is paramount, with the device built to meet strict DIN VDE and EN standards, ensuring protection against electrical hazards when used correctly. The included VoltSensor function allows for non-contact detection of phase voltages, indicated by a red LED and an acoustic signal, which is particularly useful for tracing cable breaks or identifying live phases quickly.

Powering the instrument are two integrated 1.5 V micro batteries, offering an approximate lifespan of 250 hours under normal conditions. The package includes safety measuring leads rated for CAT III 1000 V, a compact protective pouch, and a multilingual operating manual available on an included CD. The meter automatically powers off after ten minutes of inactivity to conserve energy. Its compact dimensions of 190 x 60 x 40 mm and lightweight design make it portable for field work, while the open fork opening of 16 mm accommodates single conductors effectively. Regular calibration is recommended annually to maintain accuracy, and the device must be handled with care to avoid damage from moisture or extreme temperatures.

M: QW-35-FL

© Copyright 2026 Manual.ly