Technical Data:
Measurement range: 0 – 300 mmHg
Accuracy of Pressure indication: ± 3 mmHg
Storing conditions:
–30°C – 70°C, 15 – 85% rel. humidity
Operating conditions:
10°C – 40°C, 15 – 85% rel. humidity
Cleaning of Unit and Cuff:
– To clean the unit only use a dry, soft cloth.
– To clean the cuff – remove the rubber bag. The exterior
of the cuff can then be washed by hand at a maximum
temperature of 30 degrees Celsius.
Disinfection:
For disinfectant wipes (at least 5 minutes exposure time) of
the device and the cuff, we recommend the disinfectant mi-
krozid sensitiv liquid (Schülke & Mayr).
To disinfect the cuff, we recommend spray disinfection.
Warranty Conditions:
We give 2 years warranty from the date of purchase. The
purchase date has to be proven by the invoice. Within the
warranty period defects are eliminated free of charge. After
repairs the warranty period is not extended on the whole
unit, but only to the replaced components.
Excluded from the warranty are parts subject to normal
wear and tear (e.g. cuff), transport damages and any da-
mage caused by improper handling (e.g. non-compliance
with the instructions for use). Damages due to disassembly
by unauthorized persons are also excluded from warranty.
No claims for damages against us are substantiated by the
warranty.
In the case of justified warranty claims the device has to be
sent along with the original invoice to:
BOSCH + SOHN GMBH U. CO. KG,
Bahnhofstr. 64, 72417 Jungingen,
GERMANY.
0124
Handmod_solid_5spr_1612sd.qxd:Handmod_solid_5-spr_703 14.12.2016 10:13 Uhr Seite 7

Chat with your Boso Roid II manual using AI

Get instant AI answers from your Boso Roid II manual

Ask anything about the Boso Roid II. Answers come from this manual, with the pages they are based on.

Roid II Maintenance Schedule

Remind me when maintenance is due
Every 2 Years
Perform calibration testing

Have the device calibrated by the manufacturer or an authorized service to ensure accuracy and compliance.

As Needed
Clean the unit with a dry cloth

Use only a soft, dry cloth to clean the main body of the blood pressure monitor. Avoid moisture.

As Needed
Wash the cuff cover by hand

Remove the rubber bag from the cuff. Wash the outer cover by hand at a maximum temperature of 30 degrees Celsius.

As Needed
Disinfect the unit and cuff

Wipe the device with mikrozid sensitiv liquid for at least 5 minutes. Use spray disinfection for the cuff.

Free maintenance reminders

Timed to your purchase date. Always free, unsubscribe anytime.

By submitting you agree to the privacy policy. Unsubscribe with one click in any email.
M: qwen/qwen3.8-flash

Ask a Question

0/1000

Do you recommend Roid II?

Yes No

Frequently Asked Questions

A

Do not inflate the cuff above 300 mmHg. Exceeding this limit can damage the device and pose safety risks to the patient.

Is this answer helpful?

A

Technical calibration testing must be performed at least every two years by the manufacturer or an authorized service provider.

Is this answer helpful?

A

For children aged 3 to 12, Phase IV (K4) is recommended. This is the point where sounds change from clear to muffled.

Is this answer helpful?

A

No. Only the cuff cover can be washed by hand at 30°C. Remove the rubber bladder first. Wipe the main unit with a dry cloth.

Is this answer helpful?

A

Bosch + Sohn provides a two-year factory warranty from the date of purchase, provided you have the original invoice.

Is this answer helpful?

A

Yes, the device is marked shock-protected, meaning it has been tested to function correctly after being dropped from up to one meter.

Is this answer helpful?

M: QW-35-FL

Advantages and Disadvantages

Advantages

  • Complies with CE marking and ISO 81060-1 international standards for medical devices
  • Shock-protected design tested to withstand drops from one meter without damage
  • Adjustable air release valve allows precise control over deflation speed for accuracy
  • Covers all Korotkoff phases (K4/K5) suitable for adults, children, and pregnant women
  • Includes detailed multilingual instructions covering setup, usage, and maintenance
VS

Disadvantages

  • Requires manual auscultation skills and a separate stethoscope for operation
  • Cuff bladder must be removed before washing, adding extra steps to cleaning
  • Warranty does not extend to normal wear and tear items like the cuff itself
  • Calibration testing is mandatory every two years to maintain regulatory compliance
  • Strict temperature limits for storage and operation may restrict use in extreme environments
M: QW-35-FL

Product Specification

Operating Mode

No Battery

Measurement Type

Upper Arm

Cuff Circumference

8.66-12.60 in

Memory Function

No

Display

No

M: QW-35-FL

Product Description

Boso Roid II Blood Pressure Monitor: A Comprehensive Guide to Accurate Health Tracking

The Boso Roid II Blutdruckmessgerät represents a pinnacle of manual blood pressure measurement technology, designed for both clinical settings and dedicated home health monitoring. Manufactured by Bosch + Sohn GmbH, this device adheres to strict European medical regulations and the international ISO 81060-1 standard, ensuring reliability and precision in every reading. The kit includes the sphygmomanometer unit, an arm cuff suitable for various arm circumferences, a user manual, and an optional carrying case.

Understanding how to use the Boso Roid II Blutdruckmessgerät Manual is essential for obtaining accurate results. The device utilizes the auscultatory method, requiring the user to listen for Korotkoff sounds through a stethoscope. For adults, Phase V (K5) is recommended as the endpoint where sounds disappear, while children aged 3 to 12 should use Phase IV (K4), marking the change from clear to muffled sounds. Pregnant women typically follow the adult protocol unless specific audible noises require switching to Phase IV.

Proper usage involves significant preparation. Patients must sit comfortably with their back supported, legs uncrossed, and arms resting at heart level. Before the first measurement, a five-minute rest period is crucial to stabilize the body. During inflation, the cuff should be pumped above systolic pressure, but never exceeding 300 mmHg. The WHO recommends a deflation speed of 2 to 3 mmHg per second, adjustable via the valve screw on the device. After measurement, fully opening the release valve ensures rapid deflation.

Durability is a key feature of the Roid II, which is shock-protected and tested to withstand drops from one meter. Cleaning is straightforward; the unit requires only a dry soft cloth, while the cuff cover can be hand-washed at temperatures up to 30°C after removing the rubber bladder. Disinfection protocols are also clearly defined, recommending specific agents like Mikrozid Sensitiv Liquid for wiping surfaces. With a two-year warranty covering manufacturing defects, this monitor stands as a robust tool for maintaining cardiovascular health.

M: QW-35-FL

© Copyright 2026 Manual.ly