7c5b3cc0c6109b803902b6971fc61c50fc6d page 17
23
tion/inhalation solutions) for high-risk groups
(e.g. patients with cystic fibrosis).
•
Ensure thorough drying after each cleaning
and disinfection process. Residual moisture
or wetness can represent an increased risk of
bacterial growth.
Preparation
•
Immediately after each treatment, all parts of
the atomiser
8
and the rest of the atomiser set
used must be cleaned of medication residues
and contamination.
•
To do this, dismantle the atomiser
8
into its
individual parts.
•
Remove the mouthpiece
12
, mask
13
14
or
nosepiece
21
from the atomiser.
•
Disassemble the nosepiece if you have used it
with the comfort attachment.
•
Dismantle the atomiser by twisting the top an-
ticlockwise against the medication container
11
.
•
Remove the atomiser insert
9
from the medi-
cation container
11
.
•
Remove the valve
10
by pulling it out of the
atomiser
8
.
•
Reassembly is carried out in reverse order.
Cleaning
Important
The device must be switched o, disconnected
from the mains and allowed to cool down each
time before cleaning.
The atomiser and the used atomiser set, such as
the mouthpiece, masks, atomiser insert, nasal
douche, etc. must be washed with hot but not
boiling water after each use. Dry the parts careful
-
ly using a soft cloth. Put the parts together again
when they are completely dry and place them in a
dry, sealed container or disinfect them.
When cleaning, ensure that any residue is re
-
moved. Never use any substances for cleaning
that could potentially be toxic if they came into
contact with the skin or mucous membranes, or if
they were swallowed or inhaled.
Use a soft, dry cloth and non-abrasive cleaning
products to clean the device.
Do not use any abrasive cleaning products and
never submerge the device in water.
Important
•
Make sure that no water gets inside the device!
•
Do not clean the device or atomiser set in the
dishwasher!
•
Do not touch the device with wet hands when
it is plugged in and do not allow water to spray
on the device. Only operate the device if it is
completely dry.
•
Do not spray any liquid into the vents! If liquid
penetrates the device this could cause dam
-
age to the electrics or other nebuliser parts and
lead to a malfunction.
Condensation, hose care
Condensation may form in the hose depending on
the ambient conditions. It is essential to remove
the moisture to prevent bacterial growth and en
-
sure proper treatment. To do so, proceed as fol-
lows:
•
Disconnect the hose
7
from the atomiser
8
.
•
The hose must remain in the nebuliser
4
.
•
Operate the nebuliser until the moisture is re-
moved by the air passing through.
•
In the event of heavy contamination, replace
the hose.
Disinfection
Please carefully observe the points below when
disinfecting your atomiser and the atomiser set.
We recommend disinfecting the individual parts
on a daily basis after the last usage as a minimum
measure.
•
First, clean the atomiser and atomiser set as
described in the “Cleaning” section.
•
Place the disassembled atomiser
8
, the nasal
douche (n/a), the mouthpiece
12
, the silicone
masks
13
14
and the disassembled nose-
piece in boiling water for 5 minutes.
•
For disinfection immerse the compressed air
hose in a 2 % sodium hypochlorite (NaOCl)
solution for 15 minutes (i.e. a solution made
from the disinfecting agent Amuchina
®
or a
2 % sodium hypochlorite solution mixed by
your pharmacist). Afterwards rinse all parts
thoroughly with tap water, to fully remove any
possible residue of the disinfectant solution.

Chat with your Beurer IH 24 Kids manual using AI

Get instant AI answers from your Beurer IH 24 Kids manual

Ask anything about the Beurer IH 24 Kids. Answers come from this manual, with the pages they are based on.

IH 24 Kids Maintenance Schedule

Remind me when maintenance is due
Daily (after last use)
Disinfect the atomiser and set

Place disassembled atomiser, masks, mouthpiece, and nosepiece in boiling water for 5 minutes. Immerse hose in 2% NaOCl for 15 mins.

Show in the manual (page 23)
Yearly or every 200 hours
Replace air filter

Under normal conditions, replace the air filter after approximately 200 operating hours or one year.

Show in the manual (page 22)
Yearly
Replace atomiser and set

Replace the atomiser and rest of the set no later than one year (assuming max 360 disinfections) due to wear affecting therapy efficiency.

Show in the manual (page 24)
Yearly
Replace nasal douche

Replace the nasal douche accessory after one year of use.

Show in the manual (page 22)
After every treatment
Clean the atomiser and set parts

Wash all parts of the atomiser and used set with hot, non-boiling water immediately after each treatment to remove residues.

Show in the manual (page 23)
Every 10-12 uses
Check air filter condition

Inspect the air filter regularly after 10-12 nebulisation procedures. Replace if dirty, clogged, or damp.

Show in the manual (page 22)

Free maintenance reminders

Timed to your purchase date. Always free, unsubscribe anytime.

By submitting you agree to the privacy policy. Unsubscribe with one click in any email.
M: qwen/qwen3.8-flash

Ask a Question

0/1000

Do you recommend IH 24 Kids?

Yes No

Frequently Asked Questions

A

It is suitable for children from 6 months under supervision and self-use from 12 years. Adults can use it too.

Is this answer helpful?

A

It treats upper and lower respiratory diseases like asthma, bronchitis, and sinusitis via aerosol therapy.

Is this answer helpful?

A

Wash parts with hot water after use. Disinfect masks and cups in boiling water for 5 minutes.

Is this answer helpful?

A

No, essential oils are unsuitable and may clog the system or cause bronchospasms.

Is this answer helpful?

A

Replace the air filter after 200 hours of use or once a year, whichever comes first.

Is this answer helpful?

A

Yes, provided you follow the 30-minute on and 30-minute off cycle to prevent overheating.

Is this answer helpful?

A

Stop usage immediately and contact customer service or an authorized retailer for repairs or support.

Is this answer helpful?

M: QW-35-FL

Advantages and Disadvantages

Advantages

  • Suitable for children aged 6 months and older with adult supervision.
  • Quiet operation at 52 dB minimizes disturbance during treatment sessions.
  • Includes a storage bag for organized travel and home use accessories.
  • Safety features include automatic thermal protection for motor health.
  • Designed specifically for domestic use with clear user guidance.
VS

Disadvantages

  • Requires a mandatory 30-minute rest period after 30 minutes of operation.
  • Filter component must be replaced annually or after 200 operating hours.
  • Not intended for professional clinical settings or hospital environments.
  • Cleaning process is mandatory after every single use session.
  • Essential oils are not compatible and may damage the system.
M: QW-35-FL

Product Specification

Type

Compressor

Operation Mode

Mains powered

Mist Output

0.01 fl oz/min

Features

Suitable for children, disinfectable, medicine containers

Package Contents

Bag, mouthpiece, childrens mask, baby mask, elbow piece, filter set, connecting hose

M: QW-35-FL

Product Description

Beurer IH 24 Kids Inhalator Review

Respiratory Health for Families

Managing breathing issues in children requires the right tools to ensure comfort and safety during daily care routines. The Beurer IH 24 Kids provides a reliable solution for home therapy within domestic environments. This compressor nebulizer helps treat upper and lower airway conditions effectively without invasive procedures. Parents often look for safety and ease of use in medical devices to manage respiratory health. This model meets those needs with specific design features for younger users who need gentle treatment. It targets conditions like asthma, bronchitis, and sinusitis commonly affecting young lungs. The aerosol therapy accelerates healing and prevents disease spread through medication delivery. By using prescribed medication, you minimize systemic diffusion and reduce risks elsewhere in the body. The device humidifies dry airways and loosens stubborn secretions for easier coughing. It helps relieve swelling and inflammation in the bronchial mucosa during flare-ups. Children breathe easier and sleep better during treatment sessions at home.

Operation and Safety Standards

Safety is paramount when treating kids at home with medical equipment. The device includes a 30-minute ON and 30-minute OFF cycle to prevent overheating the motor. It operates quietly at around 52 dB(A) to avoid disturbing sleeping children during nap time. Parents should always follow the Beurer IH 24 Kids Manual for proper setup and usage guidelines. The box contains a nebulizer cup, hose, mouthpiece, and silicone masks for different ages. A storage bag keeps accessories organized when traveling or moving between rooms. Cleaning is essential for hygiene to prevent bacterial growth on plastic parts. Parts must be washed with hot water after every single use cycle. Some components require disinfection in boiling water to kill germs effectively. Filters change annually or after 200 hours of actual use time. Do not use essential oils as they may clog the internal nozzle system. Consult your pharmacist regarding compatibility with silicone and PP materials found in the set.

Maintenance and Long Term Use

Keeping the device clean ensures consistent performance for every therapy session. Do not use abrasive cleaners or submerge the main electrical unit in water. Water damage can cause electrical faults and void safety certifications completely. Always unplug before cleaning the air hose or nebulizer components. The air filter protects the motor from dust entering the internal mechanism. Replace it if dirty or damp to ensure proper airflow. Do not repair the device yourself as this voids the warranty coverage. If issues arise, contact customer service or an authorized retailer for help. This product follows European medical device standards for safety and efficacy. It is not for professional clinical use in hospitals or clinics. Store it in a dry place away from children and pets. Always consult a doctor before starting therapy to confirm diagnosis. Proper usage ensures better health outcomes and faster recovery times. Regular disinfection prevents bacterial growth which could harm the patient. Use a 2 percent sodium hypochlorite solution for the hose tube. Ensure all parts dry completely before storage in the bag. Avoid direct sunlight on the storage bag to prevent plastic degradation.