Solo Maintenance Schedule
Sanitise the washed pump set parts once per day by boiling for at least 5 minutes or using microwave steam bags.
Disassemble and wash all parts that come into contact with breast milk immediately after each use using warm soapy water.
Check the tubing before use. If condensation is present, wash and dry it. If mold is found, discontinue use and replace the tubing.
Inspect pump set components for signs of wear or damage before assembly. Replace defective or worn parts if necessary.
Ask a Question
Do you recommend Solo?
Yes NoDo you recommend Solo?
Yes NoFrequently Asked Questions
It is a personal-use electric breast pump for lactating women who want to express and collect milk from one breast at a time. It is intended for a single user.
Is this answer helpful?
The pump starts with a faster stimulation phase intended to encourage milk flow. It then changes to a slower expression phase. You can press the let-down button when milk begins to flow, or wait for the automatic switch after one minute.
Is this answer helpful?
Yes. The pump can operate while connected to power, although the battery charges more slowly while the pump is running. A full charge takes about two hours.
Is this answer helpful?
Wash parts that contact breast milk after every use and sanitise them once daily. Disassemble the breast shield, connector, membrane, bottle, lid, and stand. Wipe the pump unit only; do not immerse it or pour liquid onto it.
Is this answer helpful?
Normally no, because the tubing carries air rather than milk. Inspect it regularly. If condensation appears, wash and dry it completely. If milk is visible in the tubing, contact Medela rather than attempting to wash or sanitise it.
Is this answer helpful?
Lower the vacuum to a comfortable setting and check that all parts are clean, dry, correctly connected, and sealed around the breast. If pain, minimal milk output, or weak suction continues, consult a healthcare professional, lactation specialist, or Medela customer service.
Is this answer helpful?
Advantages and Disadvantages
Advantages
- 2-Phase Expression technology uses stimulation and expression rhythms to support a natural pumping routine.
- Adjustable vacuum controls help users find their own maximum comfortable setting.
- Rechargeable battery can be used while charging, offering flexibility for daily sessions.
- Rotatable PersonalFit Flex shield and optional sizes support more personalized positioning.
- Pause function, status light, bottle stand, and automatic shutoff add practical convenience.
Disadvantages
- Designed only for single pumping, so it may not suit users seeking simultaneous double pumping.
- Included breast shield is 24 mm; other sizes require a separate purchase.
- Tube and pump unit need careful moisture management and cannot be treated like washable milk-contact parts.
- This is a single-user product, limiting sharing or hand-me-down use.
- The pump cannot be self-repaired; persistent faults require Medela customer service.
Product Specification
Accessories
1x 5.07 fl oz bottle, 1x bottle holder, 1x Personal Fit Flex connector, 1x 0.94 in Personal Fit Flex breast shield, 1x USB charging cable, 1x USB power adapter
Product Description
Medela Solo Breast Pump - Read, Download & Print Free
The Medela Solo Breast Pump is a personal-use electric pump designed for breastfeeding mothers who want a straightforward way to express and collect milk. Built for single pumping, it combines a compact pump unit, PersonalFit Flex breast shield, connector, bottle, tubing, and power adapter in one practical system. Its rechargeable battery can be charged in around two hours, and the pump may also be used while charging.
Comfortable, responsive pumping
A key feature is Medela's 2-Phase Expression technology. The pump begins with a quicker stimulation rhythm intended to encourage milk flow, then moves into a slower expression rhythm. Users can switch phases with the let-down button or allow the pump to change automatically after one minute. Adjustable vacuum controls make it easier to find a setting that feels comfortable rather than overly intense.
The included oval PersonalFit Flex breast shield can rotate 360 degrees, helping users position it for a more natural fit. Correct sizing remains important: the shield should be slightly larger than the nipple diameter. Additional shield sizes are available separately, allowing mothers to choose a fit that supports comfort and effective pumping.
Designed for everyday routines
The Solo is made for individual use and has an operating life of 250 hours. Its simple controls include buttons to increase or decrease vacuum, pause a session, change pumping phases, and turn the unit on or off. A status light indicates whether the pump is in stimulation, expression, or pause mode, while also communicating battery level.
For added convenience, the pump pauses when requested and switches off automatically after two minutes of inactivity during a pause. It also shuts down after 30 minutes when there is no user interaction. The bottle stand helps keep expressed milk stable after a session, and the bottle lid allows for secure storage preparation.
Cleaning and milk handling essentials
Parts exposed to breast milk should be washed after every use and sanitised once daily. The breast shield, connector, membrane, bottle, lid, and stand can be disassembled for cleaning. The tubing generally carries only air and does not need routine washing, but it should be inspected for condensation, milk, or mould. All components must be completely dry before assembly and use.
Freshly expressed milk can be stored at room temperature, refrigerated, or frozen according to appropriate safety guidance. Avoid microwaving or boiling breast milk, as this may create hot spots and affect its composition. For confidence and comfort, use only compatible Medela parts, maintain a clean setup, and seek advice from a healthcare professional or lactation specialist if pumping causes pain or poor milk expression.
Notes
- This pump is intended for one user only.
- Keep wireless devices at least 30 cm away during operation.
- Replace damaged, worn, mouldy, or contaminated parts promptly.