Bosotherm Medical Maintenance Schedule
For medical use, perform regular checks on the accuracy of the measurement system at least once a year by authorized personnel.
Use a new, undamaged disposable probe cover for every measurement to prevent infection and ensure accuracy.
Wipe the casing and probe with an alcohol sponge or cotton swab moistened with 70% Isopropyl alcohol. Do not immerse in water.
Replace the CR2032 lithium battery immediately when the battery symbol appears on the display.
Free maintenance reminders
Timed to your purchase date. Always free, unsubscribe anytime.
Ask a Question
Do you recommend Bosotherm Medical?
Yes NoDo you recommend Bosotherm Medical?
Yes NoFrequently Asked Questions
The thermometer uses infrared technology to provide a temperature reading in approximately 1 second once properly positioned in the ear or when measuring a surface.
Is this answer helpful?
Yes. Always fit a new, undamaged disposable probe cover for each measurement to prevent cross-contamination and to ensure accurate readings; use original boso probe covers (Art.-Nr. 520-7-086).
Is this answer helpful?
No. The device is not waterproof. Do not immerse it in liquids. Clean the casing and probe with a 70% isopropyl alcohol swab and avoid abrasive cleaners; ensure no liquid enters the instrument.
Is this answer helpful?
With the unit off, press and hold the Start button for about 5 seconds. The current unit will flash—press Start again to toggle between °C and °F, then wait 5 seconds to return to measurement mode.
Is this answer helpful?
Low readings can result from probe misalignment, curved ear canal, earwax obstruction, damaged or missing probe covers, or measuring outside the operating ambient temperature (10–40 °C). Straighten the ear canal and use a new cover for reliable results.
Is this answer helpful?
Replace the CR2032 battery when the battery icon appears flashing before a measurement or permanently after a measurement. Open the battery cover, observe polarity, and install a new CR2032 cell.
Is this answer helpful?
Error indications like ‘L’ or ‘H’ relate to measurement or ambient temperature limits; follow the manual’s troubleshooting: check ambient conditions, battery polarity/charge, probe cover condition, and contact the manufacturer if errors persist.
Is this answer helpful?
Advantages and Disadvantages
Advantages
- 1-second infrared measurement for fast results
- Accurate tympanic readings with ±0.2 °C lab accuracy in clinical range
- Hygienic disposable probe covers reduce infection risk
- Memory recall of last reading and audible fever alarm
- Wide measurement range (0 °C to 100 °C) supports surface checks
Disadvantages
- Not waterproof—must be kept dry and cannot be immersed
- Requires correct probe-cover use; damaged covers affect accuracy
- Earwax or improper probe placement can yield low readings
- Limited by ambient operating range (10–40 °C) for accurate results
- Single CR2032 battery; replacement required after ~1000 uses
Product Specification
Type
Infrared thermometer
Operating mode
Battery operated (1x button cell)
Application
in the ear (auricular)
Units of measurement
°F, °C (switchable)
Measuring range
32°F to 212°F (surface temperature)
Accuracy
±0.36°F
Measurement time
1 s
Memory function
yes
Display
yes
Special features
Fever indicator, Surface measurement, Automatic shut-off, Disinfectable
Scope of delivery
Disposable protective cap
Product Description
boso Bosotherm medical Manual
The boso Bosotherm medical is a fast, accurate infrared ear thermometer designed for home and clinical use. Using a sensitive IR sensor and a uniquely designed probe assembly, it measures tympanic infrared emissions and converts them into reliable temperature readings within one second. The unit performs an automatic self-test each time it powers on to ensure measurement integrity and displays the last recorded value on start-up.
Compact and user-friendly, the device supports body and surface temperature measurements across a wide range (0 °C to 100.0 °C / 32.0 °F to 199.9 °F). It features a clear LCD with 0.1° resolution, an audible fever alarm (10 short beeps above 37.5 °C), and simple switching between °C and °F by holding the start button during power-off. A disposable probe cover system keeps measurements hygienic and reduces cross-contamination—original boso probe covers are recommended.
Safety and handling are emphasised: the thermometer is not waterproof, must be kept dry and stored between 10–40 °C (50–104 °F), and should not be used if the probe or housing is damaged. For best results, straighten the ear canal (pull ear up and back) to align the probe with the tympanic membrane, use a new intact probe cover for every measurement, and allow short waits between repeat measures (30 seconds after several consecutive readings). The thermometer is clinically validated when used per instructions and includes guidance on cleaning (70% isopropyl alcohol swabs), disinfecting, battery replacement (CR2032), and disposal.
Technical highlights include laboratory accuracy of ±0.2 °C within the clinical range (32.0–42.2 °C), automatic shutoff after ~1 minute, memory of the last reading, approx. 1,000 measurements per CR2032 battery, compliance with EN12470-5, and a small, lightweight design.
This versatile instrument suits parents monitoring infants, caregivers tracking fevers, and general household use for quick body or surface temperature checks. Proper technique, clean probe covers, and adherence to operating temperature ranges ensure dependable results.
Notes
- Use only original boso probe covers (Art.-Nr. 520-7-086).
- Not waterproof: never immerse in liquids.
- Replace battery when low-battery icon appears; observe polarity when installing.
- For critical patients (newborns, immunocompromised children), take three readings in the same ear and use the highest value.
- Report any unexpected operation or health-worsening incident to the manufacturer promptly.