measuring sensor display field ON/OFF-button battery compartment cover
Measuring methods
Oral
30 seconds
normal temp.:
35,7°C – 37,2°C
Rectal
10 seconds
normal temp.:
36,2°C – 37,7°C
Axillary
30 seconds
normal temp.:
35,2°C – 36,7°C
MEASURING METHODS
• In the mouth (oral):
Put the thermometer underneath your tongue. The instrument must have
good contact with the tongue. Close the mouth and breath calmly through
your nose in order not to influence the measurement by the inhaled air. Mi-
nimum measuring time: 30 seconds
• In the anus (rectal):
This is the most reliable method and the most appropriate for babies and
children. Carefully immerse the tip of the thermometer 2-3 cm into the anus.
Minimum measuring time: 10 seconds
• Underneath the arm (axillary):
Under medical aspects this measuring method will lead to
inaccurate results. Therefore it is not appropriate to get exact measure-
ments. Minimum measuring time: 30 seconds
IMPORTANT SAFETY INSTRUCTIONS
• The thermometer is appropriate for the measurement of the body tempera-
ture only!
• This device can be used at home and in hospital.
• The user should be at least 11 years old.
• The patient can be the user.
• Do not interrupt the measurement before the signalling tone [beep] indica-
tes that the minimum measurement time is completed.
• The thermometer contains small parts (battery, etc.) that can be swallowed
by children. Therefore never leave the thermometer unattended to children.
• Protect the thermometer from tremor and shock!
• Avoid bending of the tip more than 45°!
• Avoid ambient temperatures above 60°C. NEVER boil the thermometer!
• Only water or the disinfectants listed in the section «cleaning and disinfec-
ting» are to be used for cleaning. The thermometer must be intact when it is
immersed in liquid disinfectant.
• The thermometer has a long-life battery. For disposal, please comply with
local regulations for special waste.
• Regular checks on the accuracy of the measurement system should be car-
ried out in accordance with the valid directives of the individual countries,
governing this issue.
• Wireless communication devices, such as home networking devices, mo-
bile phones, cordless phones and their base stations, walkie-talkies can af-
fect this thermometer. Therefore, a minimum distance of 11 feet should be
kept from such devices.
10 sec.
30 sec.
signal tone
memory
100% waterproof (IP 67)
oral/axillary
rectal
read instructions of use
boso_bosotherm_flex_kpl_1804sd.qxd:boso_bosotherm_flex_kpl_805 05.04.2018 12:49 Uhr Seite 5

Chat with your Boso Bosotherm Flex manual using AI

Get instant AI answers from your Boso Bosotherm Flex manual

Ask anything about the Boso Bosotherm Flex. Answers come from this manual, with the pages they are based on.

Bosotherm Flex Maintenance Schedule

Remind me when maintenance is due
Every 2 years
Perform measurement accuracy check

Check the accuracy of the measurement system. This must be done by the manufacturer, the competent authority, or qualified persons.

After every use
Disinfect the device

Wipe the entire device completely with Antifect Liquid and observe a working time of approximately 5 minutes.

As needed
Replace battery

Replace the battery when the double downward arrows symbol appears on the display.

Free maintenance reminders

Timed to your purchase date. Always free, unsubscribe anytime.

By submitting you agree to the privacy policy. Unsubscribe with one click in any email.
M: QW-35-FL

Ask a Question

0/1000

Do you recommend Bosotherm Flex?

Yes No

Frequently Asked Questions

A

The manual specifies 30 seconds for oral (under the tongue) and for axillary (under the arm). Rectal measurement is listed as the fastest at a minimum of 10 seconds. Always wait until the signal tone indicates completion.

Is this answer helpful?

A

From a medical perspective, the manual states that axillary measurements can be inaccurate and therefore are not suitable for exact readings. For more reliable results, the manual describes oral and especially rectal measurement as better options.

Is this answer helpful?

A

If the temperature is below 32.0 C, the display shows "---" for low. If it is above 42.9 C, the display shows "H" for high. This helps you quickly recognize measurements outside the supported clinical range.

Is this answer helpful?

A

During measurement, the unit shows the changing temperature while the C symbol flashes. When the beeper sequence sounds and the symbol stops flashing, the manual indicates the temperature rise is below 0.1 C over 8 seconds, meaning the measurement is complete.

Is this answer helpful?

A

For cleaning, the manual instructs using 70 percent isopropyl alcohol. For disinfection, wipe the device completely with Antifect Liquid and follow the stated contact time of about 5 minutes. The thermometer should be intact and protected from damage.

Is this answer helpful?

A

When the LCD shows the "▼▼" symbol, the battery is depleted and should be replaced. Open the battery compartment cover and insert a new battery with the plus (+) symbol facing upward, then close the compartment securely.

Is this answer helpful?

M: OA-GPT-54-NAN

Advantages and Disadvantages

Advantages

  • Supports oral, rectal, and axillary measurements with stated minimum times
  • Beeper and stabilization rule help confirm when the measurement is complete
  • Clear LCD plus range indicators (low "---", high "H")
  • Internal self-test at 37.0 C and automatic memory of the last result
  • 100 percent waterproof design rated IP67 for easier cleaning and disinfection
VS

Disadvantages

  • Axillary method is stated to yield less accurate results and is not recommended for exact measurements
  • Fixed clinical temperature range (32.0 C to 42.9 C) with warnings outside it
  • Requires careful handling to avoid bending the probe beyond 45 degrees
  • Disinfection and cleaning must follow specific products and time guidance
  • Wireless devices can interfere, so adequate separation is recommended
M: OA-GPT-54-NAN

Product Specification

Type

digital stick thermometer

Usage

in the mouth (oral), in the armpit (axillary), in the rectum (rectal)

Measurement units

°F

Measurement range

89.6°F to 109.22°F (body temperature)

Accuracy

±10.18°F

Measurement time

10s

Memory function

yes

Display

yes

Special features

flexible tip, waterproof

M: GO-GEM-FL-25

Product Description

Boso Bosotherm flex probe thermometer Manual - Reliable body temperature measurement

The Boso Bosotherm flex probe thermometer is a clinical maximum thermometer designed for at-home and hospital use. It uses a flexible probe so you can take readings in the mouth (oral), the rectum (rectal), or under the arm (axillary). The display shows results with a clear LCD readout and signals when the minimum measuring time has been reached.

This thermometer supports typical measurement ranges between 32.0 C and 42.9 C. If the temperature is too low, the display shows "---"; if it is too high, it shows "H". Accuracy is listed as plus or minus 0.1 C within the clinical range, and the device runs an internal self-test at 37.0 C to help confirm correct operation.

For measurement timing, it follows simple guidance: oral measurements take at least 30 seconds, rectal measurements take at least 10 seconds, and axillary measurements take at least 30 seconds. The manual notes that the axillary method can be less accurate from a medical perspective, so it is not intended for the most exact readings.

Comfort and readiness are supported by a beeper that confirms when the thermometer is on and when the temperature rise has stabilized (the rise is described as less than 0.1 C over 8 seconds). After use, you can extend battery life by switching the unit off briefly; otherwise it turns off automatically after about 10 minutes. The thermometer also stores the last measurement so you can quickly review it.

Because hygiene and practical care matter, the probe is designed for wet cleaning and disinfection. The manual specifies cleaning with 70 percent isopropyl alcohol, and disinfection using Antifect Liquid with an indicated contact time of about 5 minutes. The device is described as 100 percent waterproof with IP67 protection.

Notes

  • Rectal measurement is described as the most reliable method, especially for infants and young children.
  • Do not interrupt measurements before the signal tone indicates the minimum time is complete.
  • Keep the device away from strong impacts and avoid bending the probe beyond 45 degrees.

Key results (pros)

  • Multiple measurement sites: oral, rectal, and axillary with clear minimum times
  • Beep guidance and stability criterion help indicate when readings are complete
  • LCD with 0.1 C resolution and range warnings for low and high temperatures
  • Automatic storage of the last value plus a built-in self-test
  • IP67 waterproof design suitable for cleaning and disinfection

Key results (cons)

  • Axillary readings are explicitly described as potentially inaccurate and not ideal
  • Measurement range is limited to 32.0 C to 42.9 C
  • Includes small parts (battery) that should be kept away from children
  • Care instructions are specific (70 percent alcohol and Antifect Liquid)
  • Wireless communication equipment may affect readings if distance is too short
M: OA-GPT-54-NAN